#1 The Problems of “Creeping mAs/Dose” in America 01/01/13
As this is my first blog, I decided to
jump right in and tackle what I believe is one of the biggest problems in
radiology today; and that is the over radiation that is occurring everywhere with digital
x-ray (CR and DR). As of January 1st, 2013, well over half of the radiology departments in the US are now digital.
The first CR equipment came out in 1983,
which means it has now been around for 30
years. That's thirty years!! The problem is most of us
never had proper training when it came to the new techniques which implements
using at least 15% more kV and cutting the mAs in half. By doing this the patient’s entrance dose is
cut by 33%. Most vendors either had facilities
utilize the same techniques that were used with film, a slightly higher kV with
the same or even more mAs.
Most training consisted of being shown how
to correctly set up and get an exposure and post process the image, and that
was about it. The whole concept of a
“light” image on your monitor no longer exists but if the patient is under
radiated the image could have mottle/noise.
To take no chances, many radiographers slowly started to increase the
mAs little by little to guarantee a perfect “looking” image every time. The
problem is the computer has the ability to automatically rescale the image and
make it look perfect even though 5, 10, 50
times too much mas/dose/radiation is
used. The Exposure Index (EI) numbers
show that the patient has been over radiated but these numbers are not utilized
nearly enough (see my next blog on the EI numbers later this month).
To make matters even worse, students don’t
really learn to be as discriminating in determining the exact size of a patient
so that they can use the perfect technique.
Also there is so much latitude with how much mAs can be used in digital
radiography, it’s impossible for any radiographer or student to be as good with
manual techniques as we had to be with film. So each year the mAs slowly and insidiously
creeps up a little bit at a time but over a 10-20 year period departments are
now using double and triple the mAs that is needed
Finally, how do we combat this
problem?
1.
Everyone needs to
be better educated about using higher kV and lower mAs.
2. Radiographers need to teach the students how to
visually access the correct size of a patient (or use calipers).
3. Everyone needs to be better trained in using the EI
numbers.